Instrument:Gas Analyzer
Manufacture:Licor
Model No:LI-7700
Serial No:TG1-0216
Tag No:
Acquisition Date:2024-08-16
Project:Borrowed
Location on 2024-08-19:Lab,
Notes: Borrowed from AMP for Eden Landing for Kuno's chamber campaign. Bare instrument with hat - no cable. DOE # WD79219

Documents

Manual: Collecting_LI-7700_Ethernet_Data.pdf
Manual: LI-7700_Manual_V2_201103.pdf
Manual: LI-7700_Manual_V3_201312.pdf

History

Rec No.DateLoc FromTitleLoc ToNotes
6812 2024-08-19 Lab none Loaned Loaned none Given to Patty Oikawa for use at Eden Landing. Included were a yellow power cable and teal ethernet cable that belong to Biomet Lab.
6810 2024-08-16 Lab none Calibration Lab none Calibration Date:2024-08-16
Cal Location: Lab
Last calibration: 2021-12-01
Calibrated by: Joe
RSSI=69.5 Optics RH=23, Ambient P=100.4kPa, Tair=22.5C
Zero Air at 1400psi, AMP CA03923 3.3865ppm 1500psi
CH4 Offset: 0.025ppm
CH4 Offset: 0.0025ppm
Notes: Swapped desiccant bottle, new optics RH about 9%