Date: |
2013-02-21 |
From: |
Mayberry - Gas Analyzer CH4 |
To: |
Mayberry - Gas Analyzer CH4 |
Title: |
Calibration |
Notes: |
Calibration Location: Mayberry
Last calibration: 2012-09-06
RSSI=85%, Optical RH=3%
Ambient P=101.97kPa, Tair=14.4C
Zero Air at 1800psi, 2nd CH4 in Air 4.982 at 1300
CH4 Zero Offset: 0.011ppm
CH4 Span Offset: 0.010ppm
Notes: The 7700 had lost line lock and was manually lock prior to calibration.
|
Entered by: |
|
|
|