Date: |
2013-02-19 |
From: |
West Pond - Gas Analyzer CH4 |
To: |
West Pond - Gas Analyzer CH4 |
Title: |
Calibration |
Notes: |
Calibration Location: West Pond
Last calibration: 2012-09-06
RSSI=66.8%, Optical RH=1.0%
Ambient P=101.98kPa, Tair=10.8C
Zero Air at 1800psi, 2nd CH4 in Air 4.982 at 1300
CH4 Zero Offset: 0.016ppm
CH4 Span Offset: 0.006ppm
Notes:
|
Entered by: |
|
|
|