Date: |
2013-02-12 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Upgrade |
Notes: |
Returned from Licor with AIU-0674 after repair of condensation problem. The firmware was updated and it was recalibrated.
CO2: A=1.50515E2, B=4.94461E2, C=4.84055E7, D=-1.58118E10, E=2.05944E12, XS=-0.0014, Z=-2.10000E-3, zero=0.8676, span=1.0038
H2O: A=4.89101E3, B=3.36421E6, C=-5.74265E7, XS=-0.0014, Z=2.13000E-2, zero=0.8539, span=0.9906, CX=40300
SS: B=2.213, C=3.942 |
Entered by: |
|
|
|