Date: |
2013-02-11 |
From: |
Factory - |
To: |
Lab - none |
Title: |
Upgrade |
Notes: |
Returned from Licor with AIU-0397 after repair of condensation problem. The firmware was updated and it was recalibrated.
CO2: A=1.59506E2, B=1.38372E3, C=5.57490E7, D=--1.93676E10, E=2.63113E12, XS=-0.0002, Z=-4.20000E-3, zero=0.8837, span=1.0037
H2O: A=5.37667E3, B=4.39566E6, C=-3.30591E8, XS=-0.0013, Z=1.62000E-2, zero=0.8417, span=0.9978, CX=41300
SS: B=2.154, C=4.196 |
Entered by: |
|
|
|