Equipment and Calibration History

View Equipment |

View history entry 885 for Los Gatos Research FMA RMT-200 sn: 08-0114

Date: 2013-01-29
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2012-8-29 Cell Pressure=142.76Torr, Cell Temperature=21.69C, Ring Time=11.939usec 2nd CH4 in air at 1400psi CH4 Span=4.982ppm CH4 = 4.959 -> 4.982 Offset: 23.000000000001ppb Notes: Calibrated to check new 400ppm CO2 secondary standard. I took and average of 75 points as recorded in the LGR internal files and got a concentration of 3.856ppm CH4
Entered by: