Equipment and Calibration History

View Equipment |

View history entry 88 for Licor LI-7500A sn: 0073

Date: 2010-03-18
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Calibrated in the lab 100.46kPa, 17.08C, 58% AGC 12.8C Td N2 tank 0.0 @ 700psi, CO2 tank 597.4ppm @ 200psi Last cal 11-4-2009 CO2 zero 0.9587 -> 0.9576 offset -6.4ppm CO2 span 0.9973 -> 0.9960 offset +0.6ppm H2O zero 0.7325 -> 0.7346 offset +0.238g/m3 H2O span 0.9865 -> 1.0007 offset -0.2C SDM address 7
Entered by: