Date: |
2012-12-21 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Returned |
Notes: |
Returned from Licor with AIU-0666 after repair of condesation problem. Condesation was removed from the source lens and the detector was noisy so was also replaced. The 7500 was recalibrated.
CO2: A=1.57076E2, B=5.35482E3, C=5.64100E7, D=-1.93457E10, E=2.76431E12, XS=0.0025, Z=-9.00000E-4, zero=0.8901, span=1.0010
H2O: A=4.99824E3, B=6.61401E6, C=-1.74913E9, XS=-0.0007, Z=1.06000E-2, zero=0.7016, span=0.9935, CX=32000
SS: B=2.691, C=4.178 |