Date: |
2012-12-21 |
From: |
Factory - |
To: |
Lab - none |
Title: |
Returned |
Notes: |
Returned from Licor with AIU-0650 after repair of condensation problem. The source lense assembly was replaced and it was recalibrated.
CO2: A=1.60238E2, B=4.98171E3, C=5.27446E7, D=-1.82900E10, E=2.48774E12, XS=0.0034, Z=-2.90000E-3, zero=0.9145, span=1.0090
H2O: A=5.66730E3, B=4.24355E6, C=-3.61850E8, XS=-0.0011, Z=2.16000E-2, zero=0.8788, span=0.9989, CX=42900
SS: B=5.060, C=5.350 |