Date: |
2012-09-06 |
From: |
Mayberry - Gas Analyzer CH4 |
To: |
Mayberry - Gas Analyzer CH4 |
Title: |
Calibration |
Notes: |
Calibration Location: Mayberry
Last calibration: 2012-01-19
RSSI=58%, Optical RH=%
Ambient P=101.37kPa, Tair=17.0C
ultra zero air at 1900psi, ch4 in air 4.982 at 950
CH4 Zero Offset: 0.034ppm
CH4 Span Offset: 0.016ppm
Notes: Calibrated in the field by Joe and Cove
|
Entered by: |
|
|
|