Equipment and Calibration History

View Equipment |

View history entry 8 for Vaisala GMP221 sn: Y0710013

Date: 2009-12-01
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Vaisala factory calibration 0-3% - certificate on file. Problem: Probe label is 3%. probe is spanned to 2% and is not scaled correctly. Action: Re-spanned probe to 3%, udated coefficients and adjusted unit
Entered by: