View history entry 771 for Licor LI-7500A sn: 75H-2176

Date: 2012-08-29
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Calibration Date:2012-08-29 Last calibration: 2012-01-27 AGC=44 SDM=6 DAC1= DAC2= Ambient P=100.32kPa Tair=26.67C, Tdew=13.58C Zero air at 1800psi, CO2 in air 398.9ppm 2000psi Tdewspan=15.08 CO2 zero 0.8691 -> 0.8694 Offset: 1.4ppm CO2 span 1.0076 -> 1.0152 Offset: -3.8ppm H2O zero 0.8509 -> 0.8534 Offset: 14.3mmol/m3 H2O span 0.9947 -> 0.9979 Offset: 0C Notes: This IRGA seemed to be reading low in the field but no major offset seen in calibration. First lab calibration since receiving this instrument.