Equipment and Calibration History

View Equipment |

View history entry 7268 for Licor LI-7500A sn: 75H-2669

Date: 2025-06-09
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2025-01-27 AGC=98.8 SDM=7 DAC1=- DAC2=- AIU sn AIU-2652, Presure A0=59.516, A1=15.503 Ambient P=100.19kPa, Tair=21.6C, Tdew=12.4C Zero Air at 1900psi, AMP CA04396 417.61 at 1250 Tdewspan=16.0 CO2 zero 0.9186 -> 0.9185 Offset: -0.47ppm CO2 span 1.0152 -> 1.0170 Offset: -0.91ppm H2O zero 0.9018 -> 0.9022 Offset: 2.4ppm H2O span 1.0215 -> 1.0206 Offset: 0.01ppm Notes:
Entered by: