View history entry 7224 for Licor LI-7500A sn: 75H-2667

Date: 2025-05-08
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2025-01-07 AGC=99.5 SDM=7 DAC1=- DAC2=- AIU sn AIU-2652, Presure A0=59.516, A1=15.503 Ambient P=101.0kPa, Tair=19.1C, Tdew=9.8C Zero Air at 100psi, AMP CA03923 488.75 at 1200 Tdewspan=13.0 CO2 zero 0.941 -> 0.9418 Offset: ppm CO2 span 1.0165 -> 1.0218 Offset: ppm H2O zero 0.8454 -> 0.8421 Offset: ppm H2O span 1.0393 -> 1.0332 Offset: ppm Notes: I started to calibrate this head yesterday but the H2O zero offset was large ~-30 and opposite for the third time. I decided to swap the internal chemicals - they were hard and needed a screw driver to pry out of the bottles. I tried to flush the body with zero air but the zeros were still high ~18ppm CO2. I let it run overnight. The chopper is a little noisy and I had to use 0.1ppm less to get the CO2 span to read right.