Equipment and Calibration History

View Equipment |

View history entry 722 for Licor LI-7500A sn: 0042

Date: 2012-08-02
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2012-05-17 AGC=53 SDM=7 DAC1=0-1000 DAC2=10-25 Ambient P=100.49kPa, Tair=21.28C, Tdew=11.87C Zero Air at 1900psi, CO2 in air 436.53 at 1400 Tdewspan=15.08 CO2 zero 0.9546 -> 0.9513 Offset: -12.72ppm CO2 span 1.0182 -> 1.0044 Offset: 7.97ppm H2O zero 0.8346 -> 0.8472 Offset: 72.5ppm H2O span 0.8284 -> 0.9696 Offset: -3.2ppm Notes: Relatively large offsets seem to counter similar large offsets seen at last calibration. H2O zero offset may be a typo.
Entered by: