View history entry 7149 for YSI EXO fDOM sn: 22C100780

Date: 2025-02-27
From: Lab - none
To: Lab - none
Title: Calibration
Notes: fDOM (QSU and RFU), Last cal: 2025-02-27 EXO sn 23H107001 EXO/Sensor Firmware: 1.0.86/3.0.0 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 QSU,-0.09 QSU,0.01 QSU,0.00 QSU,23.418 °C Pt2,302.7 QSU,307.42 QSU,302.70 QSU,0.00 QSU,18.463 °C ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 RFU,0.02 RFU,0.00 RFU,0.00 RFU,23.198 °C Pt2,100.90 RFU,100.60 RFU,100.90 RFU,0.00 RFU,19.207 °C Note: Made fresh calibration solution yesterday using QSE solution from 2024-07-20 but then forgot to update temperature compensated standard value. Stored solution in fridge overnight and let warm up to room temp (3hrs) today before calibration
Entered by: Daphne Szutu