View history entry 7145 for YSI EXO Turbidity sn: 24F102410

Date: 2025-02-27
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Turbidity (FNU), Last cal: 2024-10-22 EXO sn 23H107001 EXO/Sensor Firmware: 1.0.86/3.0.5 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 FNU,-0.04 FNU,0.00 FNU,0.00 FNU,23.311 °C Pt2,124.00 FNU,114.35 FNU,124.01 FNU,0.00 FNU,23.413 °C Pt3,1010.00 FNU, 931.74 FNU, 1009.99 FNU, 0.00 FNU, 23.782 °C Note: Much larger offset for 1010 FNU point that we have seen before, brand new cal solutions
Entered by: Daphne Szutu