Equipment and Calibration History

View Equipment |

View history entry 7133 for Licor LI-7700 sn: TG1-0423

Date: 2025-02-19
From: East End -
To: East End -
Title: Calibration
Notes: Calibration Location: East End Last calibration: 2024-08-07 RSSI=76%, Optical RH=7% Ambient P=102.2kPa, Tair=14.5C Zero Air at 450psi, AMP CA03923 3.3865 at 1300 CH4 Zero Offset: 0.01ppm CH4 Span Offset: -0.0135ppm Notes: Starting to rain but hood kept mirrors clean during calibration
Entered by:
Field notes