Equipment and Calibration History

View Equipment |

View history entry 7110 for Licor LI-7500A sn: 75H-2669

Date: 2025-01-27
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2024-10-16 AGC=99.2 SDM= DAC1= DAC2= AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=100.7kPa, Tair=23.3C, Tdew=1.84C Zero Air at 825psi, AMP FB04319 514.07 at 300 Tdewspan=20.00 CO2 zero 0.9203 -> .9186 Offset: -6.74ppm CO2 span 1.0143 -> 1.0152 Offset: -.72ppm H2O zero 0.8979 -> .9018 Offset: 24.8ppm H2O span 1.0193 -> 1.0215 Offset: -.05ppm Notes: Software span point of 514.07 --> CO2 reading of 514.14 Software span point of 514.02 --> CO2 reading of 514.07
Entered by: