Equipment and Calibration History

View Equipment |

View history entry 7082 for Licor LI-7500A sn: 75H-2667

Date: 2025-01-07
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2024-07-19 AGC=98.8 SDM= DAC1= DAC2= AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=101.0kPa, Tair=25.9C, Tdew=2.99C Zero Air at 900psi, AMP FB04319 514.07 at 400 Tdewspan=20.00 CO2 zero 0.9416 -> .9410 Offset: -2.32ppm CO2 span 1.0177 -> 1.0165 Offset: .64ppm H2O zero 0.8407 -> .8454 Offset: 29.2ppm H2O span 1.0371 -> 1.0393 Offset: -.04ppm Notes: Larger H2O zero offset (29.2 mmol/m3) than usual, but it was offset by the offset from the previous calibration (-20.2 mmol/m3). There is a consistent drift in the H2O zero coefficient since 2023. I don't think the chemicals have been swapped since 2016, so I could do that next time. Used "514.03" as CO2 span point in the 7550 software.
Entered by: