Equipment and Calibration History

View Equipment |

View history entry 7048 for YSI EXO Wiped Cond/Temp sn: 24F100050

Date: 2024-07-25
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Sp Cond (uS/cm),Last cal: EXO sn 23H107001 EXO/Sensor Firmware 1.0.86/3.0.5 CellConstant 0.47 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,5000.0 µS/cm,4954.6 µS/cm,5000.0 µS/cm,0.0 µS/cm,22.816 °C Note: First calibration on new sensor. Used USGS standard from Fondriest that expired Aug 2023. It was stored in fridge w/ cap taped and I brought it close to room temperature before calibration.
Entered by: Daphne Szutu