Date: |
2024-07-25 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Sp Cond (uS/cm),Last cal:
EXO sn 23H107001
EXO/Sensor Firmware 1.0.86/3.0.5
CellConstant 0.47
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,5000.0 µS/cm,4954.6 µS/cm,5000.0 µS/cm,0.0 µS/cm,22.816 °C
Note: First calibration on new sensor. Used USGS standard from Fondriest that expired Aug 2023. It was stored in fridge w/ cap taped and I brought it close to room temperature before calibration.
|
Entered by: |
Daphne Szutu |
|
|