Equipment and Calibration History

View Equipment |

View history entry 7042 for YSI EXO Turbidity sn: 24F102410

Date: 2024-10-22
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Turbidity (FNU), Last cal: 2024-07-25 EXO sn 23H107001 EXO/Sensor Firmware: 1.0.86/3.0.5 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 FNU,-0.06 FNU,0.00 FNU,0.00 FNU,23.324 °C Pt2,124.00 FNU,123.12 FNU,124.00 FNU,0.00 FNU,22.559 °C Pt3,1010.00 FNU, 1004.19 FNU, 1010.01 FNU, 0.00 FNU, 22.612 °C Note:
Entered by: Daphne Szutu