View history entry 7040 for YSI EXO Turbidity sn: 22E106072

Date: 2024-10-22
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Turbidity (FNU), Last cal: 2024-07-21 EXO sn 22C103416 EXO/Sensor Firmware: 1.0.86/3.0.4 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 FNU,0.33 FNU,0.00 FNU,0.00 FNU,23.484 °C Pt2,124.00 FNU,140.26 FNU,124.00 FNU,0.00 FNU,22.445 °C Pt3,1010.00 FNU, 1133.83 FNU, 1010.01 FNU, 0.00 FNU, 22.568 °C Note:
Entered by: Daphne Szutu