Equipment and Calibration History

View Equipment |

View history entry 7039 for YSI EXO fDOM sn: 24F104438

Date: 2024-10-22
From: Lab - none
To: Lab - none
Title: Calibration
Notes: fDOM (QSU and RFU), Last cal: 2024-07-22 EXO sn 22C103416 EXO/Sensor Firmware: 1.0.86/3.0.0 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 QSU,0.10 QSU,0.00 QSU,0.00 QSU,23.760 °C Pt2,300.00 QSU,319.62 QSU,300.01 QSU,0.00 QSU,22.150 °C Notes: Calibration solution from 2024-07-21. First calibration on new sensor. ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 RFU,-.02 RFU,0.00 RFU,0.00 RFU,23.208 °C Pt2,100.0 RFU,106.05 RFU,100.00 RFU,0.00 RFU,22.127 °C Notes: Cal solution using QSE solution from 2024-07-20.
Entered by: Daphne Szutu