Equipment and Calibration History

View Equipment |

View history entry 7038 for YSI EXO fDOM sn: 22C100780

Date: 2024-10-22
From: Lab - none
To: Lab - none
Title: Calibration
Notes: fDOM (QSU and RFU), Last cal: EXO sn 23H107001 EXO/Sensor Firmware: 1.0.86/3.0.0 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 QSU,-1.75 QSU,0.00 QSU,0.00 QSU,23.325 °C Pt2,300.0 QSU,272.10 QSU,300.00 QSU,0.00 QSU,22.418 °C ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 RFU,-0.61 RFU,0.00 RFU,0.00 RFU,23.221 °C Pt2,100.00 RFU,91.28 RFU,100.00 RFU,0.00 RFU,22.440 °C Note: Made fresh calibration solution using QSE solution from 2024-07-20. Calibrated with a mess-up calibration solution yesterday so did a factory reset before today's calibration.
Entered by: Daphne Szutu