Date: |
2024-10-22 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
fDOM (QSU and RFU), Last cal:
EXO sn 23H107001
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 QSU,-1.75 QSU,0.00 QSU,0.00 QSU,23.325 °C
Pt2,300.0 QSU,272.10 QSU,300.00 QSU,0.00 QSU,22.418 °C
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 RFU,-0.61 RFU,0.00 RFU,0.00 RFU,23.221 °C
Pt2,100.00 RFU,91.28 RFU,100.00 RFU,0.00 RFU,22.440 °C
Note: Made fresh calibration solution using QSE solution from 2024-07-20. Calibrated with a mess-up calibration solution yesterday so did a factory reset before today's calibration.
|
Entered by: |
Daphne Szutu |
|
|