Equipment and Calibration History

View Equipment |

View history entry 7012 for Campbell CR1000 sn: 45312

Date: 2024-11-08
From: Lab - none
To: Lab - none
Title: Upgrade
Notes: I am having trouble getting FTP and ping to work on this logger. I've ruled out problems with the NL115/116, cables and ethernet switch. I could connect to the CR1000 with an IP port. It was running OS v32.02 and the newest OS is v32.07. I updated the OS hoping it would wipe whatever setting might be interfering. Nope - no luck. All the hardware works just fine on another CR1000.
Entered by: Joe Verfaillie