Equipment and Calibration History

View Equipment |

View history entry 6975 for Licor LI-7500A sn: 75H-2514

Date: 2024-10-22
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2024-07-23 AGC=99.9 SDM= DAC1= DAC2= AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=100.6kPa, Tair=25.4C, Tdew=11.9C Zero Air at 1100psi, AMP FB04319 514.07 at 850 Tdewspan=20.00 CO2 zero 0.9086 -> .9089 Offset: .9ppm CO2 span 0.988 -> .9962 Offset: -5.71ppm H2O zero 0.8457 -> .8476 Offset: 12.5ppm H2O span 0.9823 -> .9936 Offset: -.26ppm Notes: