View history entry 6973 for Licor LI-7500A sn: 0041

Date: 2024-10-22
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2024-06-17 AGC=98.9 SDM= DAC1= DAC2= AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=100.7kPa, Tair=25.7C, Tdew=11.4C Zero Air at 1050psi, AMP FB04319 514.07 at 800 Tdewspan=20.00 CO2 zero 0.9958 -> .9960 Offset: .54ppm CO2 span 1.0176 -> 1.0142 Offset: 2.36ppm H2O zero 0.8238 -> .8244 Offset: 3.3ppm H2O span 1.0341 -> 1.031 Offset: .08ppm Notes: