View history entry 6901 for Licor LI-7700 sn: TG1-0514

Date: 2024-10-07
From: Lab -
To: Lab -
Title: Calibration
Notes: Calibration Location: Lab Last calibration: 2021-07-12 RSSI=79%, Optical RH=21% Ambient P=100.2kPa, Tair=31.5C Zero Air at 1400psi, AMP FB04035 1.19156 at 1000 CH4 Zero Offset: -0.057ppm CH4 Span Offset: 0.0704ppm Notes: First check of instrument from USGS. Last zero was 2023-09-08. Clock 25min behind, set heater and wash to Biomet Default. TC was dead, swapped with repaired one. Initial span used ppb instead of ppm so off by a factor of 1000. Redid span on 10/08 with different tank - seemed fine less than ppb off.