Equipment and Calibration History

View Equipment |

View history entry 6854 for Licor LI-7500A sn: 1594

Date: 2024-09-26
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2024-01-10 AGC=99.8 SDM=0 DAC1=- DAC2=- AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=100.1kPa, Tair=23.7C, Tdew=14.0C Zero Air at 1350psi, AMP CA03923 488.75 at 1450 Tdewspan=20.0 CO2 zero 0.8774 -> 0.8769 Offset: -2.0ppm CO2 span 1.0093 -> 1.0086 Offset: 0.4ppm H2O zero 0.8142 -> 0.8147 Offset: 3.0ppm H2O span 1.0062 -> 1.0096 Offset: -0.08ppm Notes:
Entered by: