Date: |
2012-07-27 |
From: |
Factory - |
To: |
Lab - none |
Title: |
Repair |
Notes: |
Returned from Licor. Under warrenty they replaced the spar and spar clamps. After repair it would lose signal strength under cold conditions. Replacing the spinning mirror assembly fixed this problem. It was fully calibrated:
Zero = -23.6232, span = 9.5816e-05, lasermoddepth = 17000, laserstarttemp = 21, blockstarttemp = 30, blockstarttemplowrange = 5, rssidropthresh = 0, pzero = 57.5, pspan = 15.27, samplegain = 0, refgain = 0, mirrorpos = 0, offset1 = 32000, delta1 = 0, offset2 = 0, delta2 = 0, dither = 30000, sampledcoffset = 443, sampleacoffset = -965, sampleopticaloffset = 0 |