Equipment and Calibration History

View Equipment |

View history entry 685 for Licor LI-7700 sn: TG1-0155

Date: 2012-07-27
From: Factory -
To: Lab - none
Title: Repair
Notes: Returned from Licor. Under warrenty they replaced the spar and spar clamps. After repair it would lose signal strength under cold conditions. Replacing the spinning mirror assembly fixed this problem. It was fully calibrated: Zero = -23.6232, span = 9.5816e-05, lasermoddepth = 17000, laserstarttemp = 21, blockstarttemp = 30, blockstarttemplowrange = 5, rssidropthresh = 0, pzero = 57.5, pspan = 15.27, samplegain = 0, refgain = 0, mirrorpos = 0, offset1 = 32000, delta1 = 0, offset2 = 0, delta2 = 0, dither = 30000, sampledcoffset = 443, sampleacoffset = -965, sampleopticaloffset = 0
Entered by: