View history entry 6836 for Licor LI-7500A sn: 75H-2150

Date: 2024-09-03
From: Lab - none
To: Loaned - none
Title: Calibration
Notes: Last calibration: 2024-06-14 AGC=103 SDM= DAC1= DAC2= AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=100.21kPa, Tair=26.14C, Tdew=14.36C Zero Air at 1350psi, AMP CA03923 488.75 at 1450 Tdewspan=20.00 CO2 zero 0.9036 -> .9026 Offset: -4.89ppm CO2 span 1.026 -> 1.0162 Offset: 6.1ppm H2O zero 0.8737 -> .8716 Offset: -15.1ppm H2O span 1.0443 -> 1.0188 Offset: .57ppm Notes: Larger offsets than usual but I think it's ok because it's the first calibration after swapping internal chemicals and running the sensor on the bench for 5 days. Used 488.7 as the span point in the software. Installed back on BWW tower after calibration.