View history entry 6816 for Licor LI-7810 sn: TG10-01751

Date: 2024-08-12
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last zero 2023-12-01 Zero Air, 1450psi CH4 zero offset: 3.4ppb CO2 zero offset: -2.8ppm Last span 2023-12-01 AMP CA03923, 3386.5ppb CH4, 488.75ppm CO2, 1500psi CH4 span offset: -1.25ppb CO2 span offset: 0.96 ppm Previous weight: 0.9980 New weight: ?? Notes: Calibration in the lab before Kuno's field campaign next week. Ran zero and span gases at our usual flow rate of 2 SCFH and plumbed the system with the recommended 30cm vent before the air inlet to vent excess air pressure and to prevent diffusion of ambient air into the gas stream. I spoke with a Licor technical person at AGU (Graham I think) who said that because the laser is tuned to methane, we only need to wait for CH4 to stabilize for the calibration (don't need to wait for H2O).
Entered by: Daphne Szutu