View history entry 6805 for Licor LI-7700 sn: TG1-0223

Date: 2024-08-07
From: Gilbert Tract -
To: Gilbert Tract -
Title: Calibration
Notes: Calibration Location: Gilbert Tract Last calibration: 2024-01-11 RSSI=85%, Optical RH=7% Ambient P=100.5kPa, Tair=30.7C Zero Air at 1600psi, AMP CA03923 3.3865 at 1800 CH4 Zero Offset: -.011ppm CH4 Span Offset: -.0085ppm Notes: Warm, breezy