Date: |
2024-07-24 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
fDOM (QSU and RFU), Last cal: 10/27/2023 14:54 AM
EXO sn 23H107001
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 QSU,-2.37 QSU,0.00 QSU,0.00 QSU,23.604 °C
Pt2,313.80 QSU,290.85 QSU,313.83 QSU,0.00 QSU,12.006 °C
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 RFU,-0.84 RFU,0.00 RFU,0.00 RFU,24.327 °C
Pt2,104.60 RFU,91.47 RFU,104.61 RFU,0.00 RFU,12.002 °C
Note: Used calibration solution from 2024-07-21. This sensor was returned to us from USGS on 2024-07-23 at Gilbert Tract.
|
Entered by: |
Daphne Szutu |
|
|