Calibration |
Notes: |
fDOM (QSU and RFU), Last cal:
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 QSU,-1.64 QSU,0.00 QSU,0.00 QSU,22.291 °C
Pt2,310.80 QSU,283.56 QSU,310.81 QSU,0.00 QSU,14.025 °C
Notes: Calibration solution from 2024-07-21. First calibration on new sensor.
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 RFU,-.53 RFU,0.00 RFU,0.00 RFU,22.076 °C
Pt2,104.1 RFU,94.92 RFU,104.10 RFU,0.00 RFU,13.123 °C
Notes: Calibration solution from 2024-07-21. First calibration on new sensor.
|