Equipment and Calibration History

View Equipment |

View history entry 6787 for YSI EXO Turbidity sn: 22E106072

Date: 2024-07-21
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Turbidity (FNU), Last cal: 2024-04-08 EXO sn 22C103416 EXO/Sensor Firmware: 1.0.86/3.0.4 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 FNU,-1.05 FNU,0.00 FNU,0.00 FNU,21.060 °C Pt2,124.00 FNU,116.00 FNU,124.00 FNU,0.00 FNU,21.062 °C Pt3,1010.00 FNU, 926.15 FNU, 1009.99 FNU, 0.00 FNU, 21.104 °C Note: 0 FNU point done with MQ water (stored in plastic bottle at room temeprature since 2024-02-01)
Entered by: Daphne Szutu