Equipment and Calibration History

View Equipment |

View history entry 6786 for YSI EXO fDOM sn: 22E100649

Date: 2024-07-21
From: Lab - none
To: Lab - none
Title: Calibration
Notes: fDOM (QSU and RFU), Last cal: 2024-04-08 EXO sn 22C103416 EXO/Sensor Firmware: 1.0.86/3.0.0 ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 QSU,-0.72 QSU,-0.01 QSU,0.00 QSU,22.62 °C Pt2,300.00 QSU,164.59 QSU,300.03 QSU,0.00 QSU,21.826 °C Notes: Fresh calib solution but sensor is way off. Probably send back to factory. ,Standard,PreCal,PostCal,RawValue,Temp Pt1,0.00 RFU,-.33 RFU,0.00 RFU,0.00 RFU,21.804 °C Pt2,100.0 RFU,RFU,34.39 RFU,100.00 RFU,0.00 RFU,21.927 °C Notes: Fresh calib solution but sensor is way off. Probably send back to factory.
Entered by: Daphne Szutu