Equipment and Calibration History
View Equipment |
Date: |
2024-03-06 |
From: |
Factory - none |
To: |
Lab - none |
Title: |
|
Notes: |
Received from YSI as a warranty replacement for sn 22E103340. Came with an extra brush and a small allen key for installing the brush. The factory inspection reported that water had leaked into the probe housing and all internal components were corroded. They replaced the wiper under warranty. YSI recommends sending the wiper back to the factory every year to have wiper motor o-rings replaced and lubed to prevent water leakage. |
Entered by: |
Daphne Szutu |
|
|