Equipment and Calibration History

View Equipment |

View history entry 6759 for Licor LI-7500A sn: 75H-2514

Date: 2024-07-22
From: Lab - none
To: Lab - none
Title: Calibration
Notes: Last calibration: 2024-05-06 AGC=97.8 SDM= DAC1= DAC2= AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=99.79kPa, Tair=22.1C, Tdew=13.18C Zero Air at 1650psi, AMP CA04119 524.77 at 350 Tdewspan=20.00 CO2 zero 0.91 -> .9086 Offset: -5.85ppm CO2 span 1.0009 -> .9880 Offset: 9.33ppm H2O zero 0.846 -> .8457 Offset: -2.0ppm H2O span 0.9936 -> .9823 Offset: .26ppm Notes: Larger CO2 zero (-5.85ppm offset) and CO2 span (9.33ppm offset) than expected. Checked CO2 span again after calibrating everything else and it read 524.52ppm (offset=-0.25ppm). That looks ok so I left the calibration as is.
Entered by: