Equipment and Calibration History

View Equipment |

View history entry 6756 for Licor LI-7500A sn: 75H-2176

Date: 2024-07-21
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2024-07-19 AGC=10.3 SDM= DAC1= DAC2= AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=100.0kPa, Tair=22.17C, Tdew=13.43C Zero Air at 1675psi, AMP CA04119 524.77 at 350 Tdewspan=20.00 CO2 zero 0.8677 -> .8676 Offset: -.39ppm CO2 span 1.0132 -> 1.0116 Offset: 2.03ppm H2O zero 0.8578 -> .8573 Offset: -2.7ppm H2O span 0.9377 -> .9357 Offset: .04ppm Notes: Second calibration since internal chemical swap on 2024-07-15. Looks good. Used "524.75" as the CO2 span point in the 7550 software.
Entered by: