Equipment and Calibration History

View Equipment |

View history entry 6755 for Licor LI-7500A sn: 75H-2176

Date: 2024-07-19
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2024-03-23 AGC=103 SDM= DAC1= DAC2= AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=100.4kPa, Tair=21.2C, Tdew=16.5C Zero Air at 1700psi, AMP CA04119 524.77 at 350 Tdewspan=20.00 CO2 zero 0.8682 -> .8677 Offset: -2.2ppm CO2 span 1.0079 -> 1.0132 Offset: -3.94ppm H2O zero 0.8494 -> .8578 Offset: 52.3ppm H2O span 0.9236 -> .9377 Offset: -.34ppm Notes: First calibration after swapping internal chemicals on 2024-07-15. It has been running in the lab since the chemical swap. I used "524.75" as the CO2 span point in the 7550 software. I will test it again in a few days to make sure it's stable.
Entered by: