Equipment and Calibration History

View Equipment |

View history entry 6752 for Licor LI-7500A sn: 0042

Date: 2024-07-12
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2024-03-23 AGC=101.5 SDM= DAC1= DAC2= AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=99.9kPa, Tair=23.17C, Tdew=13.9C Zero Air at 1750psi, AMP CA04119 524.77 at 350 Tdewspan=20.00 CO2 zero 0.9483 -> .9482 Offset: -.53ppm CO2 span 1.0098 -> 1.0114 Offset: -1.07ppm H2O zero 0.8451 -> .8455 Offset: 2.1ppm H2O span 1.02 -> 1.0264 Offset: -.14ppm Notes:
Entered by: