Equipment and Calibration History

View Equipment |

View history entry 6738 for Licor LI-7500A sn: 75H-2182

Date: 2024-06-17
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2024-01-10 AGC=100.1 SDM=0 DAC1=- DAC2=- AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=100.1kPa, Tair=20.7C, Tdew=9.1C Zero Air at 1750psi, AMP CA04119 524.77 at 350 Tdewspan=61.0 CO2 zero 0.885 -> 0.8850 Offset: 0.10ppm CO2 span 1.0163 -> 1.0141 Offset: 1.57ppm H2O zero 0.8326 -> 0.8332 Offset: 3.9ppm H2O span 1.0352 -> 1.0365 Offset: -0.02ppm Notes:
Entered by: