Equipment and Calibration History

View Equipment |

View history entry 6737 for Licor LI-7500A sn: 0041

Date: 2024-06-17
From: Lab -
To: Lab -
Title: Calibration
Notes: Last calibration: 2024-01-03 AGC=100.0 SDM=0 DAC1=- DAC2=- AIU sn AIU-1866, Presure A0=58.31, A1=15.48 Ambient P=100.1kPa, Tair=20.5C, Tdew=8.8C Zero Air at 1750psi, AMP CA04119 524.77 at 400 Tdewspan=16.0 CO2 zero 0.9963 -> 0.9958 Offset: -1.60ppm CO2 span 1.0217 -> 1.0176 Offset: 2.93ppm H2O zero 0.8245 -> 0.8238 Offset: -3.9ppm H2O span 1.0365 -> 1.0341 Offset: 0.06ppm Notes: Reset SS from 101.5
Entered by: