2024-04-08 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Turbidity (FNU), Last cal: 2024-01-22
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.4
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 FNU,-2.02 FNU,0.00 FNU,0.00 FNU,22.248 °C
Pt2,124.00 FNU,122.66 FNU,124.00 FNU,0.00 FNU,21.856 °C
Pt3,1010.00 FNU, 1024.22 FNU, 1010.01 FNU, 0.00 FNU, 21.956 °C
Note: 0 FNU point done with MQ water (stored in plastic bottle at room temeprature since 2024-02-01). I also tested it with DI water from the lab tap; pre-cal value was -2.14 FNU, almost the same as the MQ water. In the future I think using DI water from the lab tap is good enough for the calibration.
|