Title: |
Calibration |
Notes: |
fDOM (QSU and RFU), Last cal: 2024-01-23
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.0
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 QSU,-0.54 QSU,0.00 QSU,0.00 QSU,22.695 °C
Pt2,300.00 QSU,132.47 QSU,300.02 QSU,0.00 QSU,22.144 °C
Notes: Fresh calibration solution but pre-cal value is still way off. Maybe because of a dirty/cloudy sensor. This sensor probably needs to go back to the factory.
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 RFU,-.31 RFU,0.00 RFU,0.00 RFU,22.796 °C
Pt2,100.0 RFU,RFU,41.74 RFU,100.00 RFU,0.00 RFU,22.184 °C
Notes: Fresh calibration solution but pre-cal value is still way off. Maybe because of a dirty/cloudy sensor. This sensor probably needs to go back to the factory.
|