Date: |
2024-02-26 |
From: |
Lab - |
To: |
Lab - |
Title: |
Calibration |
Notes: |
Last calibration: 2023-11-09
AGC=102.8 SDM= DAC1= DAC2=
AIU sn AIU-1866, Presure A0=58.31, A1=15.48
Ambient P=100.4kPa, Tair=24.1C, Tdew=10.96C
Zero Air at 350psi, AMP CA04119 575 at 524.77 Tdewspan=20.00
CO2 zero 0.9041 -> .9030 Offset: -5.59ppm
CO2 span 1.018 -> 1.0175 Offset: .48ppm
H2O zero 0.8731 -> .8737 Offset: 3.5ppm
H2O span 1.0265 -> 1.0382 Offset: -.25ppm
Notes: BWW tower sensor. One of the mount screws wouldn't come out initially; Joe got it out pulsing with vice grips + WD-40. He cleaned out the white dust inside the bolt holes with a thread cutter. He sprayed the bolts and bolt holes with a anti-corrosion spray before I installed it again. The 7550 temperature sensor was bad. I replaced with a repaired homemade temperature sensor. The HMP fan was not spinning, but it seemed to work fine after I removed a bit of fluff.
|
Entered by: |
|
|
|