Date: |
2024-02-16 |
From: |
Lab - |
To: |
Lab - |
Title: |
Calibration |
Notes: |
Calibration Location: Lab
Last calibration: 2023-02-02
RSSI=81%, Optical RH=12%
Ambient P=100.98kPa, Tair=21.5C
Zero Air at 450psi, AMP CA04119 4.2238 at 675
CH4 Zero Offset: .032ppm
CH4 Span Offset: .0058ppm
Notes: Calibration before loaning this sensor to J.J. to use at Elkhorn Slough Yampah tower while they're waiting for replacement/repair
|
Entered by: |
|
|
|