Date: |
2024-01-22 |
From: |
Lab - none |
To: |
Lab - none |
Title: |
Calibration |
Notes: |
Turbidity (FNU), Last cal: 2023-10-16
EXO sn 22C103416
EXO/Sensor Firmware: 1.0.86/3.0.4
,Standard,PreCal,PostCal,RawValue,Temp
Pt1,0.00 FNU,3.85 FNU,0.00 FNU,0.00 FNU,21.571 °C
Pt2,124.00 FNU,122.47 FNU,124.00 FNU,0.00 FNU,20.745 °C
Pt3,1010.00 FNU, 959.99 FNU, 1010.00 FNU, 0.00 FNU, 20.651 °C
Note: Used the lab's DI water for zero point for this and previous calibrations. I'll try to get some miliQ water for future calibrations since this zero point had a larger than usual offset. Probably doesn't affect our data too much since our turbidity is usually >50 FNU. |
Entered by: |
Daphne Szutu |
|
|